FujiFilm DR-ID 1300 Service Manual page 228

Digital radiography
Table of Contents

Advertisement

(3) Click [GET].
The SE board conditions will be displayed in the pop up window.
- Board Status
Acquires the board conditions in the SE, and displays the contents.
Normal:
Hardware Status : true
Configuration Status : true
Battery Status : true (when there is a battery)
MCU Status : true
SDCard Status : true
Defective: Other than above.
- Temperature
Displays the temperature of the stand SE.
Display example: Temperature = 29.4 deg C
018-234-06E
1200_400088E.ai
- SE Button Status
Displays the button status of the SE.
Normal:
Extra Sleep Button Status: true or false
Memory Image Button Status: true or false
- Battery
u NOTE u
Not used in the DR-ID 1300.
Displays the following information on the battery
Battery ID: 0 to 65535
Manufacture Date:YYYY/MM/DD
Insert Withdraw Count:0 to 65535
Temperature:0 to 65535[0.1K]
Status:0xXXXX
Cycle Count:0 to 65535
First Use:
First Use Date:YYYY/MM/DD
(4) Check the board state, and click [Cancel].
DR-ID 1300 / DR-ID 1300PU Service Manual
MU1-63
MU1-63