n Creating Calibration Data and Checking Result
(1) Click [OK].
If the data generated will be stored on the HDD, "Succeeded" will appear after it
has finished.
u NOTE u
If the data generation failed, an error number appears on the status screen.
In that case, check the position of the calibration phantom and the irradiation
field, and then perform again the "■ Exposing Calibration". If an error still occurs,
obtain the following data and report the failure.
- LogAll (IMO.log / Calib.log)
(2) Check that the calibration data have been generated.
REFERENCE
The data are stored in the following location. Check the time stamp of the file
generated.
- C:\Program Files\FujiFilm\MC\opt\config\machine\correct\PANEL2\
LPMarkerData.txt
018-234-05E
n Changing the Calibration Position and Exposing Calibration
(1) Change the setting of [Calib Position] from the value selected in the
step (6) of "■ Preparing for Calibration" to "CENTER_BOTTOM".
Then, perform "■ Preparing for Calibration", "■ Installing Calibration
Phantom", "■ Exposing Calibration", and "■ Creating Calibration Data
and Checking Result".
DR-ID 1300 / DR-ID 1300PU Service Manual
IN1-84
IN1-84