FujiFilm DR-ID 1300 Service Manual page 87

Digital radiography
Table of Contents

Advertisement

Error
Error Name
Code
11216
Battery high temperature error
Battery low temperature
11217
warning
11218
Battery low temperature error
Panel high temperature error
11219
(SE board ambient temperature)
Panel low temperature error (SE
11220
board ambient temperature)
11225
Panel self-diagnosis error
11226
Not supporting AUTO DETECT
11228
Panel cracking judgment error
018-234-06E
Occurrence Condition
[During bootup/routine processing]
The battery temperature rises to the banning-of-use level.
[During bootup/routine processing]
The battery temperature lowers to the warning level.
[During bootup/routine processing]
The battery temperature lowers to the banning-of-use level.
[During bootup/routine processing]
The board ambient temperature rises to the banning-of-use level (Approx.
60°C or higher).
<Reference>
This temperature is an estimate since it fluctuates depending on the SE
type and the external temperatures.
[During bootup/routine processing]
The board ambient temperature lowers to the banning-of-use level
(Approx. -10°C or lower).
<Reference>
This temperature is an estimate since it fluctuates depending on the SE
type and the external temperatures.
[During bootup/panel connection]
A self-diagnosis request has been made from the MC to the panel and the
response from the panel was NG.
<Reference>
The detected SE serial ID can be confirmed in the error log.
[During bootup/panel connection]
At the time of active AUTO DETECT, connection of a panel not compatible
with AUTO DETECT has been detected as the result of checking the SE
serial ID.
<Reference>
The detected SE serial ID can be confirmed in the error log.
[During menu registration]
The SE has detected a panel cracking and informs the MC.
DR-ID 1300 / DR-ID 1300PU Service Manual
Probable Cause and Remedy
- Remove the battery from the FPD. If the removed
battery remains hot, replace it with a new one, as the
battery is defective.
- Remove the battery from the FPD. If the temperature
of the removed battery lowers, check the cables in the
FPD.
- Check the ambient temperature.
- Check the ambient temperature.
- Check the ambient temperature.
- Check the thermistor (temperature sensor).
- Check the ambient temperature.
- Check the thermistor (temperature sensor).
- Execute [SE(n) Board State] of SE by MUTL and
confirm that the board has no abnormality.
- Confirm that the network between MC-SE us connected
normally.
- Inspect the boards (RMV, wireless module, POW, GTD,
ABA, ABB) and the cables in the panel.
- Check that no panel not compatible with AUTO DETECT
is connected.
- Check for wrong input of an SE serial ID with PC-TOOL.
- Panel breakage (9 mm x 9 mm or more) is detected.
MT-10
MT-10