(4) Insert the second X-Ray shot cable (optional) and the bucky cable
from the MP rear, and connect the cable terminals with the terminal
blocks (TB1 and TB2) of the AC bucky relay unit.
WARNING
Be sure to connect the X-Ray shot cable and the FG line of the bucky cable to
the FG (TB2-6) of the terminal block. If the FG line is fixed to the MP frame or the
like, electrification or ignition might result due to difference in ground levels.
u NOTE u
If the second X-Ray shot cable and the second bucky cable are not present, the
procedure is not necessary. Proceed to the procedure (5).
u NOTE u
The terminals to be connected with the second X-Ray shot cable and the second
bucky cable are indicated in white cells in the table below.
No.
TB1
X-Ray shot cable
Bucky cable
1
X-Con BKY
SHOT1 RDY1 (L)
2
X-Con BKY
SHOT1 RDY1 (N)
3
X-Con
SHOT1 RDY2 (L)
4
X-Con BKY
SHOT1 RDY2 (N)
5
BKY
SHOT1 RDY2 (L)
6
X-Con BKY
SHOT2 RDY1 (L)
7
X-Con BKY
SHOT2 RDY1 (N)
8
X-Con
SHOT2 RDY2 (L)
9
X-Con BKY
SHOT2 RDY2 (N)
10
BKY
SHOT2 RDY2 (L)
018-234-04E
TB2
X-Ray shot cable
Bucky cable
X-Con BKY
Stand AC (L)
X-Con BKY
Stand AC (N)
X-Con BKY
Bed AC (L)
X-Con BKY
Bed AC (N)
N.C.
N.C.
X-Con BKY
Stand FG
X-Con BKY
Bed FG
X-Con BKY
Stand Exp A
X-Con BKY
Stand Exp B
X-Con BKY
Bed Exp A
X-Con BKY
Bed Exp B
TB1
1
2
3
4
5
6
7
8
9
10
DR-ID 1300 / DR-ID 1300PU Service Manual
X-Ray shot cable
Bucky cable
TB2
1
2
3
4
5
6
7
8
9
10
Bucky cable
X-Ray shot cable
IN1-33
600_700051.ai
IN1-33

Hide quick links: