(11) Check that the machine-specific data is restored.
I.
Back up the CORRECT DATA on the RU-PC TOOL.
{MU2:1.13_BACKUP}
II. Open the LagData.txt with the Word Pad.
Check that the value does not remain the same, but that it gradually gets
smaller.
REFERENCE
Defect calibration results are written to "LagData.txt". Since they do not attenuate
in default, the values do not change, but if the installing of machine-specific data
succeeds, the values will gradually get smaller because the machine-specific
data that is attenuated will be replaced.
018-234-04E
10.13 Forced Transfer of the SE Correct Data
NOTE
After the machine-specific data is installed, make sure to transfer the correct data to
the panel unit where the machine-specific data have been installed. If the correct data
are not transferred, correction is not properly made, causing an artifact.
10.13.1 Forced Transfer of the SE Correct Data
(1) Start up the MUTL.
(2) Click [CALIBRATION] command and [SE Correct Data Transfer >>],
input the target SE number, and then click [SET].
#1 Click: [CALIBRATION]
#2 Click: [SE Correct Data Transfer >>]
#3 Input: Target SE number
#4 Click: [SET]
CALIBRATION window
The SE correct data stored in the MC is transferred to the SE.
NOTE
Make sure that "COMPLETE_DATASAVE" is displayed in the SE Correct Data
Status >> window. "STILL_DATASAVE" is displayed during transfers.
(3) For the other two panel units, perform step (2).
(4) Click [BACK].
DR-ID 1300 / DR-ID 1300PU Service Manual
IN2-73
#1
#2
SE Correct Data Transfer >>
window
#3
#4
1200_700185E.ai
IN2-73