(7) Click [Start].
"OK" appears in the result display area upon completion of the preparation for
exposure.
u NOTE u
If you click [Start] during automatic offset update (for 30 seconds every approx.
10 minutes), "Error 12700 currently unavailable" appears. Click [Start] again after
an interval of approx. 30 seconds, in this case.
(8) Click [Display ShotRdy] and confirm display of "SHOT_READY" in the
result indication column.
u NOTE u
If "NOT_READY" is indicated, wait for approx. 5 seconds, try again, and proceed
to the next step (exposure) after "SHOT_READY" is indicated.
When exposure is executed in "NOT_READY" status, the calibration fails and an
abnormal image is generated.
018-234-04E
n Installing the Calibration Phantom
(1) Install the included calibration phantom at the boundary between the
TOP and the CENTER panel units when selecting "Top_Center" for
[Calib Position], and when selecting "Center_Bottom", at the boundary
between the CENTER and the BOTTOM panel units.
u INSTRUCTION u
The following illustrates an example of the included calibration phantom installed
at the boundary between the TOP and the CENTER panel units. Align the
calibration phantom with the points designating the installation position printed on
the panel surface. For the panel placed vertically, attach the calibration phantom
with curing tape or other adhesives.
(A)
(B)
(A) Boundary between the TOP and the CENTER panel units
(B) Boundary between the CENTER and the BOTTOM panel units
(C) Installation positions for the calibration phantom
(D) Align two calibration phantoms on the specific installation position for each
boundary.
REFERENCE
The calibration phantom is the metal stick that is included with the MP.
\
DR-ID 1300 / DR-ID 1300PU Service Manual
(C)
(D)
1300_700023E.ai
1300_700057.ai
IN2-82