Input the serial ID mentioned on the machine-specific CD, and then
III.
click [Set].
u NOTE u
The first letter of the serial ID must be entered as a capital letter.
#1 Input: Serial ID
#2 Click: [SET]
(5) Set the IP address of the SE.
{IN1:10.10_Setting the IP Address of the SE}
{IN2:10.11_Setting the IP Address of the SE}
(6) Install the communication setting file to the panel unit whose RMV65A
board was replaced.
(7) Update SE application software version.
{IN1:10.11_Updating SE Application Software Version}
{IN2:10.10_Updating SE Application Software Version}
(8) Restore the calibration data.
{MU2:1.14_RESTORE}
(9) Perform the forced transfer of the SE correct data and check the SE
correct data save status.
{IN2:10.13_Forced Transfer of the SE Correct Data}
(10) Restart the MP.
018-234-02E
n Fuse Locations
#1
#2
1200_300069.ai
DR-ID 1300 / DR-ID 1300PU Service Manual
F2
F11
F9
F1
F10 F8
MC-24
F3
F4
F7
F6
1200_300003.ai
MC-24