Operational procedure
Step 1. Press
which you want to set the gate shape.
Step 2.
Step 3. Press
Softkey
Maximum
Normal
Wide
Minimum

Setting gate range

Specify the gate range in time. The ends of the range are defined as the
points shown in Figure 9-9. You can set the gate range by specifying the start and stop
times or the center and span. The
gate range you can set.
Lower limit
Upper limit
Operational procedure
Step 1. Press
which you want to set the gate range.
Step 2.
Step 3. Press each of the following softkeys to specify the gate range. At the side of the set value
area in the data entry bar, the distance corresponding to the set time is displayed as shown
in Figure 9-7. The displayed distance is a value that takes the velocity factor into
consideration.
Softkey
Start
Stop
Center
Span
NOTE
No Hardkey is provided for this setting. The hardkeys are dedicated to setting the sweep
*1. T
Chapter 9
Deleting Unnecessary Data in Time Domain (gating)
(or
- Press
to display the "Gating" menu.
Gating
and then select the gate shape from the following.
Shape
Function
Sets the gate shape to "maximum."
Sets the gate shape to "normal."
Sets the gate shape to "wide."
Sets the gate shape to "minimum."
E5070B/E5071B
*1
T
span
*1
T
span
(or
- Press
to display the "Gating" menu.
Gating
Function
Sets the start time
Sets the stop time
Sets the center of the gate in time
Sets the gate span in time
is the measurement range expressed in time obtained in "Measurement range" on
span
page 341.
Analysis in Time Domain (Option 010)
) and
(or
has the following limitations on the
) and
(or
) to activate a trace for
6 dB attenuation
) to activate the trace for
347