Calibration
Changing the Calibration Kit Definition
Step 3. Select
Sub Class3
Step 4. Press
TRL Line/Match - Set All - LINE >7G - Return
Step 5. Press
and check that the name you specified is selected as Cal Kit.
By assigning Match and Line 1/2 to subclass 1/2/3 respectively, you can calibrate 3
standards with different frequency bands for TRL line/match calibration.

Setting options for TRL calibration

This section describes how to set the reference impedance and measurement data of the
standard used for calculating the calibration plane for TRL calibration.
Step 1. Press
.
Step 2. Press
, and select a calibration kit.
Cal Kit
Step 3. Press
Modify Cal Kit
Step 4. Press
TRL Option
Step 5. Press
Impedance
Softkey
Line
System
Step 6. Press
Reference Plane
the reference plane.
Softkey
Thru
Reflect
190
.
.
.
to select the reference impedance.
Function
Calculates the calibration coefficient using the characteristic impedance of the
line standard as the reference impendence. The impedance of the line standard
is used as the center of the Smith chart, and the calibration coefficient is
calculated using the line standard's as the direction.
Calculates the calibration coefficient using the system impedance as the
reference impendence. Select this when the line impedance and the test port
impedance differ.
to select the measurement data of the standard used for calculating
Function
Uses the length of the thru/line standard to calculate the calibration plane.
Uses the reflection coefficient of the reflection standard to calculate the
calibration plane.
.