Procedure - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Figure 4-20
Connecting standards in full 4-port calibration

Procedure

Step 1. Press
calibration.
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. Press
Step 6. Connect an OPEN calibration standard to test port 1 (the connector to which the DUT is to
be connected).
Step 7. Press
Step 8. Disconnect the OPEN calibration standard connected in Step 6 and replace it with a
SHORT calibration standard.
Step 9. Press
Step 10. Disconnect the SHORT calibration standard connected in Step 8 and replace it with a
LOAD calibration standard.
Step 11. Press
Step 12. Repeat Step 6 to Step 11 on test port 2.
Chapter 4
or
.
.
Calibrate
.
4-Port Cal
.
Reflection
to start the calibration measurement.
Port 1 Open
to start the calibration measurement.
Port 1 Short
to start the calibration measurement.
Port 1 Load
to select the channel for which you want to perform the
Calibration
Full 4-Port Calibration
125

Advertisement