Calibration
2-port TRL calibration
Operational procedure
Figure 4-31
Connection of standards in 2-port TRL calibration
NOTE
The shapes of actual standards differ from those shown in Figure 4-31 which are just
symbols for illustration.
Step 1. Press
calibration.
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. Press
Step 6. Select the test ports for which you want to execute TRL 2-port calibration. (In the
following procedure, test ports you select here are expressed as test ports x and y.)
Softkey
1-2
1-3
1-4
2-3
2-4
3-4
NOTE
An asterisk (*) in the upper right of the softkey indicates that a calibration coefficient exists
138
or
.
.
Calibrate
.
2-Port TRL Cal
.
Select Ports
Function
Selects test ports 1 and 2.
Selects test ports 1 and 3.
Selects test ports 1 and 4.
Selects test ports 2 and 3.
Selects test ports 2 and 4.
Selects test ports 3 and 4.