Calibration
Calibration between Ports of Different Connector Types

Operating procedure

NOTE
This VBA macro changes the definition of the label of calibration kit 10 (calibration kit
corresponding to the lowest softkey) temporarily, performs calibration, and restores the
previous definition after completing calibration. Therefore, if the VBA macro is aborted
for some reason, the definition of calibration kit 10's label may be lost.
When you use this VBA macro, it is recommended that you do not use calibration kit 10. If
you are, you should back up calibration kit 10 before using the VBA macro.
To back up a calibration kit, use the VBA macro "SavRecCalKit.vba". You can download
this VBA macro from our product information Web site established for the Agilent
Technologies E5070B/E5071B.
1. Setting Stimulus Condition
Set the stimulus condition of the channel for which you will perform the calibration.
For information on the setting procedure, see Chapter 3, "Setting Measurement
Conditions."
2. Starting VBA Macro
Load the VBA project and run it.
Step 1. Press
Step 2. Press
Step 3. The Open dialog box appears. Specify the file name
"D:\Agilent\AdapterCharacterization.vba" and press
Step 4. Press
Figure 4-36
Adapter Characterization macro
3. Selecting a Channel
Select a channel (1 in Figure 4-36).
NOTE
The selected channel has no relation to the active channel.
156
.
.
Load Project
to start the macro (Figure 4-36).
.
Open
Chapter 4