Turning stimulus signal output on/off

You can turn on/off the stimulus signal output, but this will prevent you from performing
measurement. Therefore, you will not normally use this feature. This is mainly used to turn
the output back to on after it has been turned off by the power trip feature.
Follow these steps to turn the stimulus signal output on/off:
Step 1. Press
.
Step 2. Press
.
Power
Step 3. Press
. Each press toggles between on/off.
RF Out
When set to off, "RF OFF" is displayed in "4. Instrument Status Bar" on page 44.
Power trip
The power trip is a feature that the instrument uses to automatically turn off the output of
the stimulus signal to protect the instrument when a signal whose level exceeds the upper
limit is inputted to the test port.
If the power output is automatically turned off by the power trip feature, remove the cause
of the over-input and turn on the power output according to the above steps to restart the
measurement.

Setting fixed frequency at power sweep

The procedure for setting the fixed frequency (CW frequency) at the power sweep is as
follows:
Step 1. Press
or
frequency.
Step 2. Press
.
Step 3. Press
.
Power
Step 4. Press
.
CW Freq
Step 5. Enter the fixed frequency using the ENTRY block keys on the front panel.
Chapter 3
Setting Measurement Conditions
Setting Stimulus Conditions
to select the channel for which you want to set the fixed