Figure 4-16
Connecting standards in full 2-port calibration

Procedure

Step 1. Press
calibration.
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. Press
Step 6. Select the test ports on which you will perform full 2-port calibration. (In the procedure
below, the selected test ports are denoted as x and y.)
Softkey
1-2
1-3
1-4
2-3
2-4
3-4
NOTE
An asterisk (*) in the upper-right of the softkey indicates that the existing calibration
coefficient will be cleared if you select the test port and execute the acquisition of the
Chapter 4
or
.
.
Calibrate
.
2-Port Cal
.
Select Ports
Function
Selects test ports 1 and 2
Selects test ports 1 and 3
Selects test ports 1 and 4
Selects test ports 2 and 3
Selects test ports 2 and 4
Selects test ports 3 and 4
to select the channel for which you want to perform the
Calibration
Full 2-Port Calibration
119