Redefining a calibration kit

To change the definition of a calibration kit, follow the procedure below.
Defining a new calibration kit
1. Select the calibration kit to be redefined.
2. Define the type of standard. Select one from among the OPEN, SHORT, LOAD,
delay/THRU, and arbitrary impedance standards.
3. Define the standard coefficient.
4. Designate a standard class for the standard.
5. Save the data for the calibration kit that has been redefined.
Changing the pre-defined connector between male and female (e.g. OPNE (f) to OPEN
(m))
1. Select the calibration kit to be redefined.
2. Designate a standard class for the standard.
3. Save the data for the calibration kit that has been redefined.
Redefining a calibration kit
Step 1. Press
Step 2. Press
Step 3. Select the calibration kit to be redefined according to the list below.
Softkey
85033E
85033D
85052D
85032F
85032B
85036B/E
85031B
85050C/D
85052C
85038A/F/M
User
NOTE
If the names (labels) of calibration kits were changed prior to operation, the new names
will appear on the respective softkeys.
Chapter 4
.
.
Cal Kit
Function
Selects the "85033E" calibration kit
Selects the "85033D" calibration kit
Selects the "85052D" calibration kit
Selects the "85032F" calibration kit
Selects the "85032B" calibration kit
Selects the "85036B/E" calibration kit
Selects the "85031B" calibration kit
Selects the "85050C/D" calibration kit
Selects the "85052C" calibration kit
Selects the "85038A/F/M" calibration kit
Selects the fourth user-defined calibration kit
Changing the Calibration Kit Definition
Calibration
185