Turning On/Off Ripple Test And Displaying Results - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Limit Test
Ripple Test
In the first line, the channel number for the active channel at the time of file saving will be
output.
In the second line, the trace number for the active trace at the time of file saving will be
output.
The third line is a header indicating the segment items that are output from the fourth line
onward.
From the fourth line onward, the segment data are output.
Operational procedure
Follow the steps below to save/recall the ripple limit table. This operation should be done
by using the external keyboard and/or mouse.
Step 1. Display the ripple limit table.
Step 2. Press Export to CSV File from the Edit Ripple Limit menu to open the dialog box. For
information on the Save As dialog box, see the description in Figure 10-1, "Save As dialog
box," on page 355. At this time, CSV Files (with the extension *.csv) is selected as the file
type.
Step 3. Specify any folder in which you want to save the file, and enter the file name. Press
to save the ripple limit table displayed on the screen to a file.
Step 4. To recall the saved ripple limit table, press
RippleLimit menu to display the Open dialog box. For detailed information on each field
of the Open dialog box, see Figure 10-3, "Open dialog box," on page 357. At this time,
CSV Files (with the extension *.csv) is selected as the file type.
Step 5. Specify the folder that contains the file, and then select the file. Press
saved limit table on the screen.
NOTE
You can recall a limit table from a trace on any channel independently of the channel and
trace that were active when the limit table was saved to the file.

Turning on/off ripple test and displaying results

You can set the limit test ON/OFF for each trace individually.
Operational procedure
Follow the steps below to configure the on/off setting of the ripple test and the display
method.
Step 1. Press
ripple test function.
Step 2. Press
test function.
Step 3. Press
388
or
to activate the channel to which you want to apply the
or
to activate the trace to which you want to apply the ripple
to display the Analysis menu.
from the Edit
Import from CSV File
Save
to recall the
Open
Chapter 11

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5070bE5071b

Table of Contents