Example Of Defining The Trl Calibration Kit - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Calibration
Changing the Calibration Kit Definition
NOTE
Select
Step 18. Select the standards to be registered in the class from among standards numbered 1 through
21. To change the connectors between male and female (e.g. OPEN (f) to OPEN (m)),
select the appropriately labeled standards here.
Step 19. Repeat Step 17 and Step 18 until classes are defined for all test ports that need to be
redefined.
Step 20. Press
Step 21. Repeat Step 15 to Step 20 to redefine all classes that need to be modified.
Step 22. Press
Step 23. Press
on the screen.

Example of defining the TRL calibration kit

You need to enter the definition of the TRL calibration kit to perform TRL
calibration.Follow these steps to define the following calibration kit given as an example.
Procedure to define the name of the calibration kit
Step 1. Press
Step 2. Press
Step 3. Select a User kit that has not been registered.
Step 4. Press
Step 5. Press
Step 6. Type in a name you want.
Procedure to define Thru and Reflect
Step 1. Press
Step 2. Press
Step 3. Type in
Step 4. Select
Step 5. Set
Step 6. Press
188
to use the same standards for all test ports.
Set All
.
Return
.
Return
and input a new label for the calibration kit by using the keypad displayed
Label Kit
THRU (Delay 0 ps, Offset Loss 1.3 Gohm/s)
REFLECT(SHORT, Delay 0 ps)
MATCH (@0 to 2GHz)
LINE1 (50-ohm transmission line, Delay 54.0 ps @ 2G to 7GHz)
LINE2 (50-ohm transmission line, Delay 13.0 ps @ 7G to 32GHz)
.
.
Cal Kit
.
Modify Kit
.
Label Kit [User]
.
Define STDs
.
1:No Name - Label
.
THRU
-
STD Type
Delay/Thru
to 1.3Gohm/s, and
Offset Loss
to return to the Define STD menu.
Return
.
to 0.
Offset Delay
Chapter 4