Calibration
Changing the Calibration Kit Definition
Saving and loading definition file of calibration kit
You can save the definition file of the calibration kit for each standard into a file on a
storage device (hard disk drive or floppy disk drive), then recall it later to reproduce the
definition.
Save procedure
Step 1. Press
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. Specify a folder, enter a file name, and press
NOTE
The file is saved with the (.ckx) extension. Do not open and modify it directly.
NOTE
Never modify the contents (folders and files) of drives except for drive A: (floppy disk
drive) and drive D:. Modifying the contents of drives other than drive A: and drive D:
may give serous damage to the functions and performance of the E5070B/E5071B.
NOTE
When the floppy disk access lamp is lit, do not press the disk ejection button. Ejecting
a floppy disk forcefully when the lamp is lit may break the floppy disk or the disk drive.
Call procedure
Step 1. Press
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. Specify a folder, select a file name, and press
NOTE
When the floppy disk access lamp is lit, do not press the disk ejection button. Ejecting
a floppy disk forcefully when the lamp is lit may break the floppy disk or the disk drive.
192
.
, and select a calibration kit you want to save.
Cal Kit
.
Modify Cal Kit
to open the dialog box.
Export Cal Kit...
.
, and select a calibration kit you want to save.
Cal Kit
.
Modify Cal Kit
to open the dialog box.
Import Cal Kit...
.
Save
.
Open
Chapter 4