Setting display range
The E5070B/E5071B has the following limitations on the display range you can set.
Lower limit
Upper limit
The number of response points displayed on the graph is the same as the number of points
regardless of the response resolution.
Operation
Step 1. Press
which you want to set the display range.
Step 2.
Step 3. Press each of the following softkeys and then specify the display range. At the side of the
set value in the data entry bar, the distance corresponding to the set time is displayed
(Figure 9-7). The displayed distance is a value taking the velocity factor into consideration.
Softkey
Start
Stop
Center
Span
NOTE
You cannot use hardkeys to set the display. The hardkeys are dedicated to specifying the
sweep range.
Figure 9-7
Data entry bar
Enabling transformation function
Operation
Step 1. Press
which you want to use the transformation type.
Step 2.
Step 3. Press
*1. T
Chapter 9
*1
−
T
span
*1
T
span
(or
- Press
to display the "Transform" menu.
Transform
Function
Sets the start value of the display range in time
Sets the stop value of the display range in time
Sets the center value of the display range in time
Sets the span of the display range in time
(or
- Press
to display the "Transform" menu.
Transform
to enable (
) the transformation function.
Transform
ON
is the measurement range expressed in time obtained in "Measurement range" on
span
page 341.
Analysis in Time Domain (Option 010)
Transformation to time domain
) and
(or
) and
(or
) to activate the trace for
) to activate the trace for
343