Data Output

Saving and Recalling Instrument State

Saving and Recalling Instrument State
You can save the instrument state of the E5070B/E5071B into a file on mass storage (hard
disk drive or floppy disk drive) and then recall it later to reproduce that state. You can
select the stored data from the following four types.
Table 10-1
Stored data
State only
(State Only)
State and
calibration data
(State & Cal)
State and trace
(State & Trace)
State, calibration
data, and traces
(All)
In addition, the "user-preset" function is provided to allow the user to freely set up an
instrument state recalled when the preset function is executed. For more information on
how to set it up, see "Setting the user preset function" on page 449 .

Compatibility of files related to saving and recalling

The compatibility of saving/recalling the instrument state file is as follows:
Saved
352
Type
Saves the setting
by recalling it later into the E5070B/E5071B.
Saves the setting
*2
array
) to reproduce the state when it was saved by recalling it later into the
E5070B/E5071B. At this time, you can perform error correction of measured values
by using the recalled calibration data.
Saves the setting
and error-corrected memory array
recalling it later into the E5070B/E5071B. At this time, the traces are also recalled and
displayed on the screen.
Saves the setting
the state when it was saved by recalling it later into the E5070B/E5071B. At this time,
the calibration data and traces are also recalled.
*1.For information on the saved target, refer to Appendix C, "List of Default Values."
*2.For information on each array, refer to "Data Processing" on page 730.
Compatibility between different models
Files saved with the E5070B cannot be recalled with the E5071B.
Files saved with the E5070A can be recalled with the E5070B, although the
opposite is not possible.
Files saved with the E5071A can be recalled with the E5071B, although the
opposite is not possible.
E5070B
*1
E5070B
*2
E5071B
file
E5070A
E5071A
Stored data and usage
*1
of the E5070B/E5071B and reproduces the state when it was saved
*1
of the E5070B/E5071B and calibration data (calibration coefficient
*1
of the E5070B/E5071B and traces (error-corrected data array
*2
) to reproduce the state when it was saved by
*1
of the E5070B/E5071B, calibration data, and traces to reproduce
Recalling
*1
*2
E5071B
Y
N
N
Y
Y
N
N
Y
*2
E5070A
E5071A
N
N
N
N
Y
N
N
Y
Chapter 10