Measurement of Mixers
The frequency-offset function allows you to measure frequency-translating devices such as
mixers.
Measurement flow
Table 8-1 shows the measurement flow.
Table 8-1
Flow of Mixer Measurement
Item
"1. Setting
Frequency-Offset" on page
313
"2. Setting External Signal
Source" on page 316
"3. Avoid Spurious
Function" on page 320
"4. Changing Frequency
Data" on page 321
"5. Implementing Mixer
Calibration" on page 323
"6. Conversion Loss
Measurement" on page 323
1. Setting Frequency-Offset
The frequency-offset function allows you to implement the measurements in which
frequencies are different for each port, such as in mixer measurement. The frequency set
for each port can be applied to the port even if it is for stimulus or receiver.
NOTE
You can set the frequency-offset for each channel independently.
Chapter 8
Description
Sets frequency-offset
•
Procedures for setting frequency at each port
•
Procedures for enabling frequency-offset function
Sets external signal source (source of LO signal)
•
Preparations for controlling external signal source
•
Procedures for setting external signal source GPIB address
•
Procedures for selecting external signal source
•
Procedures for setting external signal source frequency
•
Procedures for setting external signal source power level
•
Procedures for controlling external signal source
Enables avoid spurious function.
•
Procedures for controlling avoid spurious function
Changes frequency data.
•
Procedures for changing frequency data
Implements mixer calibration
Measures conversion loss of mixer
•
Connection of measured mixer
•
Procedures for setting measurement parameters
•
Procedures for changing measurement data
Frequency-Offset Measurement (Option 008)
Measurement of Mixers
313