Calibration
Specifying Different Standard for Each Frequency

Disabling standard defined for a subclass

The following procedure shows how to disable a standard defined for a subclass. Note that
you cannot disable subclass 1 because at least one standard must exist. The following
procedure shows how to disable an OPEN standard of subclass 2.
Procedure
Step 1. Press
.
Step 2. Press
.
Cal Kit
Step 3. Select a calibration kit you want to use. (In this example, use the calibration kit defined in
Step 3 of "Defining different standard for each frequency band" on page 202.)
Step 4. Press
.
Modify Kit
Step 5. Press
Specify CLSs
Step 6. Press
, and select
Sub Class
Step 7. Select the standard type
Step 8. Select
.
None
Step 9. Press
.
Return
204
.
.
Sub Class 2
, press
, make settings for all ports.
Open