Calibration Types and Characteristics

Table 4-1 shows the different types of calibrations and the features of each method.
Table 4-1
Calibration Types and Characteristics
Calibration Method
Standard(s)
Used
No calibration
None
Response
•OPEN or
*1
*2
Calibration
SHORT
*3
•LOAD
•THRU
*3
•LOAD
1-Port Calibration
ECal module
(2-port/4-port)
•OPEN
•SHORT
•LOAD
Full 2-Port
ECal module
(2-port/4-port)
*1
Calibration
•OPEN
•SHORT
•LOAD
•THRU
Full 3-Port
ECal module
*5*1
*6
Calibration
(2-port
4-port)
•OPEN
•SHORT
•LOAD
•THRU
Chapter 4
Corrected Error Factor
None
Following 2 error terms:
•Reflection Tracking (Er)
*3
•Directivity (Ed)
Following 2 error terms:
•Transmission Tracking (Et)
*3
•Isolation (Ex)
Following 3 error terms:
•Directivity (Ed)
•Source Match (Es)
•Reflection Tracking (Er)
Following 12 error terms:
•Directivity (Ed1,Ed2)
*3
•Isolation (Ex21,Ex12)
•Source Match (Es1,Es2)
•Load Match (El12,El21)
•Transmission Tracking
(Et21,Et12)
•Reflection Tracking (Er1,Er2)
Following 27 error terms:
•Directivity (Ed1,Ed2,Ed3)
/
•Isolation
(Ex21,Ex31,Ex12,Ex32,Ex13,Ex2
*3
3)
•Source Match (Es1,Es2,Es3)
•Load Match
(El21,El31,El12,El32,El13,El23)
•Transmission Tracking
(Et21,Et31,Et12,Et32,Et13,Et23)
•Reflection Tracking (Er1,Er2,Er3)
Calibration Types and Characteristics
Measurement
Characteristics
Parameters
All parameters
• Low accuracy
• Calibration not required
S11
•Medium-level accuracy
(Reflection
•Quick calibration
characteristics
•Isolation calibration improves the
at 1 port)
accuracy in a reflection
measurement of a DUT with high
return loss
S21
•Medium-level accuracy
(1 direction
•Quick calibration
transmission
•Isolation calibration improves the
characteristics
accuracy in a transmission
at 2 ports)
measurement of a device with high
insertion loss
S11
•1-port measurement with the
(Reflection
highest degree of accuracy
characteristics
•Quick calibration with low chance
at 1 port)
of operator error
•Highly accurate 1-port
measurement
S11,S21,S12,S
•Highly accurate 2-port
22
*4
measurement
(All
•Quick calibration with low chance
S-parameters
of operator error
at 2 ports)
•Highly accurate 2-port
*4
measurement
S11,S21,S31,S
•Highly accurate 3-port
12,S22,S32,S1
*4
measurement
3,S23,S33
•Quick calibration with low chance
(All
of operator error
S-parameters
at 3 ports)
•Highly accurate 3-port
*4
measurement
Calibration