Evaluating Transmission Characteristics of a Front End Module

10. Executing Measurement

Step 1. Display the trigger menu.
Description
Display the trigger menu.
Step 2. Set the trigger source to "manual."
Setting Description
Set the trigger source to "manual."
Step 3. Set the trigger mode for channel 1 to "continuous."
Setting Description
Set channel 1 to the active channel.
Set the trigger mode to "continuous."
Step 4. Set the trigger mode for channels 2, 3, and 4 to "continuous" according to the same
procedure as in Step 3.
Step 5. Execute the measurement.
Setting Description
Generate a trigger event.
Step 6. Repeat the following procedure to execute auto scale for all of the channels.
Setting Description
Set the active channel.
Execute auto scale.
Step 7. After all of the procedures are completed, the screen shown in Figure 15-38 appears.