Calibration
Calibration Using 4-port ECal
Step 6. When you want to turn ON the isolation calibration, press
Step 7. Select the calibration type based on the list below.
Softkey
1-Port ECal
2-Port ECal
3-Port ECal
4-Port ECal
Thru ECal
Step 8. If you must select a port, the softkey for making this selection is displayed. Select a port
and start calibration. If you do not have to select a port, this step is skipped.
Step 9. The E5070B/E5071B detects the test ports connected to the ECal and then measurement
starts.
NOTE
If a test port to be calibrated is not connected to the ECal module, an error occurs.
136
*1
*2
*1.Options 313, 314, 413, and 414 only
*2.Options 413 and 414 only
(set to
Isolation
Function
Selects 1-port calibration
Selects full 2-port calibration
Selects full 3-port calibration
Selects full 4-port calibration
Selects THRU calibration
).
ON