Extending the Calibration Plane Using Network
De-embedding
Network de-embedding is a function for performing measurements, test port by test port,
by removing the characteristics of an arbitrary network defined by a Touchstone data file.
By removing the characteristics of the cable, test fixtures, etc. between the actual
calibration plane and the DUT, the calibration plane can be correspondingly extended. The
network de-embedding function can be used together with the port extension function (see
Figure 7-2).
Figure 7-2
Port extension and calibration plane extension using network de-embedding

Using the Network De-embedding Function

Step 1. Prepare a two-port Touchstone data file (.s2p format) corresponding to the network to be
removed.
Step 2. Press
Step 3. Press
Step 4. Press
performed.
Step 5. Press
Step 6. Using the dialog box that appears, select the Touchstone data file defining the
characteristics of the network to be removed. Once the file is selected, the selection of
Select Type
up, press
Step 7. Repeat Step 3 to Step 6 to set up the Touchstone data file for each port from which a
network is to be removed.
Step 8. Press
Step 9. Press
Step 10. If
Fixture Simulator
Chapter 7

Extending the Calibration Plane Using Network De-embedding

-
Fixture Simulator
.
Select Port
,
,
, or
to select the test port from which the network de-embedding is
1
2
3
4
.
User File
automatically changes to
.
Select Type - None
to turn the network de-embedding function
De-Embedding
.
Return
is
, press the key again to turn it
OFF
-
.
De-Embedding
. To cancel a user-defined file that has been set
User
.
ON
Fixture Simulator
.
ON