Calibration
1-Port Calibration (reflection test)

Procedure

Step 1. Press
calibration.
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. Press
Step 6. Select a test port (and corresponding S parameter) on which 1-port calibration will be
performed.
Softkey
1
2
3
4
NOTE
An asterisk (*) in the upper-right of the softkey indicates that the existing calibration
coefficient will be cleared if you select the test port and execute acquisition of the
calibration coefficient (pressing
Step 7. Connect an OPEN calibration standard to the test port (connector to which the DUT is to be
connected) selected in Step 6.
Step 8. Press
Step 9. Connect a SHORT calibration standard to the test port (connector to which the DUT is to
be connected) selected in Step 6.
Step 10. Press
Step 11. Connect a LOAD calibration standard to the test port (connector to which the DUT is to be
connected) selected in Step 6.
Step 12. Press
Step 13. Press
coefficients will be calculated and saved. The error correction function will also be
automatically enabled.
114
or
.
.
Calibrate
.
1-Port Cal
.
Select Port
Function
Selects port 1
Selects port 2
Selects port 3
Selects port 4
to start the calibration measurement.
Open
to start the calibration measurement.
Short
to start the calibration measurement.
Load
to terminate the 1-port calibration process. Upon pressing this key, calibration
Done
to select the channel for which you want to perform the
).
Done