Turning off ECal auto-detect function

The ECal module automatically detects the connection between E5070B/E5071B's test
ports and ECal module's ports. You can turn off this function to set ports manually.
NOTE
Even if the connection is wrong with the auto-detect function turned off, no error is
displayed.
Step 1. Press
auto-detect function.
Step 2. Press
Step 3. Press
Step 4. Press
Softkey
Auto
Manual
Step 5. Specify a test port of the E5070B/E5071B.
Softkey
Port 1
Port 2
Port 3
Port 4
*1.Only with Options 313, 314, 413, and 414.
*2.Only with Options 413 and 414.
Step 6. Specify an Ecal port for the port of the E5070B/E5071B you specified.
Softkey
Port A
Port B
Port C
Port D
*1.Only with Ecal modules supporting 3 ports and 4 ports.
*2.Only with Ecal modules supporting 4 ports.
Chapter 4
or
.
.
ECal
, and select
Orientation
Manual
Function
Turns on the auto-detect function.
Turns off the auto-detect function.
Function
Specifies port 1 of the E5070B/E5071B.
Specifies port 2 of the E5070B/E5071B.
*1
Specifies port 3 of the E5070B/E5071B.
Specifies port 4 of the E5070B/E5071B.
*2
Function
Specifies port A of Ecal.
Specifies port B of Ecal.
*1
Specifies port C of Ecal.
*2
Specifies port D of Ecal.
ECal (electronic calibration)
to select the channel for which you want to turn off the
.
Calibration
131