Setting a media type for the calibration kit

You can set a media type for the standard you use.
Setting procedure
Step 1. Press
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. Press
Softkey
Coaxial
Waveguide
NOTE
If you select waveguide as the media type, set the system impedance and the characteristic
impedance to 1Ω.
NOTE
Depending on the media type you use, the calculation method of the electrical delay, which
is required to correct the phase delay, differs. For more information, see "Electrical Delay"
on page 398.
Chapter 4
.
, and select a calibration kit.
Cal Kit
.
Modify Cal Kit
, and select a standard.
Define STDs
, and select a media type.
Media
Function
Selects coaxial as the media type.
Selects waveguide as the media type.