Calibration
3-port TRL calibration
coefficient (by pressing
Step 7.
to start the measurement of the calibration standard.
Thru/Line
Softkey
Port x-y Thru
Port x-z Thru
Port y-z Thru
Step 8.
to start the measurement of the calibration standard.
Reflect
Softkey
Port x Reflect
Port y Reflect
Port z Reflect
Step 9.
to start the measurement of the calibration standard.
Line/Match
Softkey
*1
x-y Line/Match
*2
x-y Fwd (Syx)
*2
x-y Rvs (Sxy)
x-z Line/Match
x-z Fwd (Szx)
x-z Rvs (Sxz)
y-z Line/Match
y-z Fwd (Szy)
y-z Rvs (Syz)
*1.This is equivalent to measuring
*2.Measuring
142
).
Done
Function
Executes thru/line calibration for test ports x and y.
Executes thru/line calibration for test ports x and z.
Executes thru/line calibration for test ports y and z.
Function
Executes reflection calibration for test port x.
Executes reflection calibration for test port y.
Executes reflection calibration for test port z.
Function
Executes line/match calibration for test ports x and y. Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
Executes line/match calibration for test ports x and y. Use this to perform
forward-direction measurement only.
Executes line/match calibration for test ports x and y. Use this to perform
reverse-direction measurement only.
Executes line/match calibration for test ports x and z. Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
Executes line/match calibration for test ports x and z. Use this to perform
forward-direction measurement only.
Executes line/match calibration for test ports x and z. Use this to perform
reverse-direction measurement only.
Executes line/match calibration for test ports y and z. Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
Executes line/match calibration for test ports y and z. Use this to perform
forward-direction measurement only.
Executes line/match calibration for test ports y and z. Use this to perform
reverse-direction measurement only.
x-y Fwd (Syx)
and
x-y Fwd (Syx)
x-y Rvs (Sxy)
and
.
x-y Rvs (Sxy)
is equivalent to measuring
x-y Line/Match
.
Chapter 4