Agilent Technologies ENA Series User Manual page 671

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Key Operation
(Continued)
Calibrate
Enhanced
Response
Select Port
Selects the transmission test (measurement of S
3-1 (S31
*1
1→3.
S11)
Selects the transmission test (measurement of S
4-1 (S41
*2
1→4.
S11)
Selects the transmission test (measurement of S
1-2 (S12 S22)
2→1.
Selects the transmission test (measurement of S
3-2 (S32
*1
2→3.
S22)
Selects the transmission test (measurement of S
4-2 (S42
*2
2→4.
S22)
Selects the transmission test (measurement of S
1-3 (S13
*1
3→1.
S33)
Selects the transmission test (measurement of S
2-3 (S23
*1
3→2.
S33)
Selects the transmission test (measurement of S
4-3 (S43
*2
3→4.
S33)
Selects the transmission test (measurement of S
1-4 (S14
*2
4→1.
S44)
Selects the transmission test (measurement of S
2-4 (S24
*2
4→2.
S44)
Selects the transmission test (measurement of S
3-4 (S34
*2
4→3.
S44)
Returns to the softkey display screen one level higher.
Cancel
Executes an OPEN calibration on the test port selected in the
Open
menu.
Port
Executes a SHORT calibration on the test port selected in the
Short
menu.
Port
Executes a LOAD calibration on the test port selected in the
Load
menu.
Port
Executes a THRU response calibration on the test ports selected in
Thru
the
Select Ports
Executes an isolation calibration on the test ports selected in the
Isolation
(Optional)
Select Ports
Terminates a calibration process and calculates the calibration
Done
coefficients.
Displays softkeys for canceling a calibration.
Cancel
Cancels a calibration process and clears the measurement data of
OK
standard.
Returns to the softkey display screen one level higher.
Cancel
Returns to the softkey display screen one level higher.
Return
Appendix D
Function
31
41
12
32
42
13
23
43
14
24
34
menu.
menu.
SCPI Command
) for test port
:SENS{1-16}:CORR:COLL:METH:ERES 3,1
) for test port
:SENS{1-16}:CORR:COLL:METH:ERES 4,1
) for test port
:SENS{1-16}:CORR:COLL:METH:ERES 1,2
) for test port
:SENS{1-16}:CORR:COLL:METH:ERES 3,2
) for test port
:SENS{1-16}:CORR:COLL:METH:ERES 4,2
) for test port
:SENS{1-16}:CORR:COLL:METH:ERES 1,3
) for test port
:SENS{1-16}:CORR:COLL:METH:ERES 2,3
) for test port
:SENS{1-16}:CORR:COLL:METH:ERES 4,3
) for test port
:SENS{1-16}:CORR:COLL:METH:ERES 1,4
) for test port
:SENS{1-16}:CORR:COLL:METH:ERES 2,4
) for test port
:SENS{1-16}:CORR:COLL:METH:ERES 3,4
:SENS{1-16}:CORR:COLL:OPEN
Select
:SENS{1-16}:CORR:COLL:SHOR
Select
:SENS{1-16}:CORR:COLL:LOAD
Select
:SENS{1-16}:CORR:COLL:THRU
:SENS{1-16}:CORR:COLL:ISOL
:SENS{1-16}:CORR:COLL:SAVE
:SENS{1-16}:CORR:COLL:CLE
Softkey Functions
Calibration Menu
665