Agilent Technologies ENA Series User Manual page 593

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Figure A-2
Turning off system error correction
NOTE
The System Correction Off button does not appear when system error correction has
already been turned off.
2-port TRL/LRM calibration operating procedure
1. Selecting Test Port
Select test ports (1 in Figure A-3).
NOTE
For more information on starting VBA macro, see "Operating Procedure" on page 586
.
Figure A-3
TRL/LRM main control panel
NOTE
The TRL/LRM calibration VBA macro automatically detects the ENA's active channel,
and calibration is performed for the active channel.
2. Defining Calibration Kit
Define the calibration kit you use.
Step 1. Press the
screen shown in Figure A-4.
Appendix A
button (2 in Figure A-3) to bring up the calibration kit definition
Define Cal kit
Manual Changes
Manual Changes
587

Advertisement

Table of Contents
loading