Figure 4-35
Connection of standards in 4-port TRL calibration

Operational procedure

NOTE
The shapes of actual standards differ from those shown in Figure 4-35 which are just
symbols for illustration.
Step 1. Press
calibration.
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. Press
Step 6.
Thru/Line
Softkey
Port 1-2 Thru
Port 1-3 Thru
Port 1-4 Thru
Port 2-3 Thru
Port 2-4 Thru
Port 3-4 Thru
Chapter 4
or
.
.
Calibrate
.
4-Port TRL Cal
.
Select Ports
to start the measurement of the calibration standard.
Function
Executes thru/line calibration for test ports 1 and 2.
Executes thru/line calibration for test ports 1 and 3.
Executes thru/line calibration for test ports 1 and 4.
Executes thru/line calibration for test ports 2 and 3.
Executes thru/line calibration for test ports 2 and 4.
Executes thru/line calibration for test ports 3 and 4.
4-port TRL calibration
to select the channel for which you want to execute
Calibration
145