Thru Response Calibration (Transmission Test); Procedure - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Calibration

THRU Response Calibration (transmission test)

THRU Response Calibration (transmission test)
In THRU response calibration, calibration data are measured by connecting a THRU
standard to the desired test port. This calibration effectively eliminates the frequency
response transmission tracking error from the test setup in a transmission test using that
port (Figure 4-7). It is also possible to carry out an isolation calibration using a LOAD
standard in the process of THRU response calibration. An isolation calibration will
eliminate isolation error (crosstalk error) from the test setup in a transmission test using
that port.
Figure 4-7
2-Port error model (THRU response)
Figure 4-8
2-Port Error model (THRU response + isolation)

Procedure

Step 1. Press
calibration.
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. Press
Step 6. Select the test ports (and corresponding S parameters) upon which a THRU response
calibration is to be performed.
Softkey
2-1 (S21)
110
or
.
.
Calibrate
.
Response (Thru)
.
Select Ports
Function
Selects test port 2 (input) and test port 1 (output); Corresponds to the
determination of S21
to select the channel for which you want to perform the
Chapter 4

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5070bE5071b

Table of Contents