Setting the Averaging Trigger Function

The averaging trigger function is used to execute the sweep the number of times specified
by the averaging factor with a single trigger when the sweep averaging function is ON.
Averaging trigger
On
Off
The averaging factor is cleared before the start of measurement.
NOTE
When the point trigger function is on, its setting has priority, and you need to generate
triggers based on "(number of measurement points)×(averaging factor)".
When the sweep averaging function is off, sweep is performed only once even if the
averaging trigger function is set to on.
The averaging trigger function is valid for all channels.Note that you can set the sweep
averaging function for each channel.

Using the averaging trigger function

1. Setting the averaging trigger function
When the sweep averaging function is on, follow these steps to set the averaging trigger
function.
Step 1. Press
Step 2. Press
Softkey
ON
OFF
2. Executing averaging measurement
Step 1. Press
Step 2. Press
executed the number of times specified by the averaging factor, and then the instrument
waits for the next trigger.
Chapter 5
Function
Performs the sweep the number of times specified by the
averaging factor with a single trigger.
Performs the sweep once with a single trigger.
.
.
Avg Trigger
Function
Sets the averaging trigger function to ON.
Sets the averaging trigger function to OFF.
.
. The averaging factor is cleared before the start of measurement, the sweep is
Single
Making Measurements
Setting the Averaging Trigger Function
249