Initializing The Limit Table - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Step 4. Press
to display the Ripple Limit menu.
Ripple Limit
Softkey
Ripple Test
Ripple Limit
Ripple Value
Ripple Value Band
Edit Ripple Limit
*1.To use the ripple test function, you must first define the ripple limit. For more on how to
define the ripple limit, see "Configuring ripple limit" on page 385.
Step 5. Press
to turn the ripple test ON. You can display the limit lines on the screen
Ripple Test
by turning on
Ripple Limit

Initializing the limit table

The following operations initialize the limit table.
At power-on
When presetting
When calling a ripple table with zero frequency bands
When
Clear Limit Table
Chapter 11
Function
Sets the ripple test ON/OFF.
Sets the ripple limit line display ON/OFF.
Sets how the ripple values are displayed. Available settings are off,
absolute value (difference between maximum and minimum values
within the band) display, and margin (difference between absolute value
of ripple and ripple limit) display.
Selects the band for which you want to display the ripple value.
Opens the ripple limit table for editing the ripple limit.
.
-
is pressed in the Edit Ripple Limit menu
OK
Limit Test
Ripple Test
*1
389