Agilent Technologies ENA Series User Manual page 596

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Manual Changes
Manual Changes
Figure A-5
Performing 2-port TRL/LRM calibration
Step 2. Press the
Step 3. Make the through condition for test ports 1 and 2 and press the
Figure A-6). The through measurement is performed, and an asterisk appears at the
1-2
Figure A-6
Example of performing 2-port TRL/LRM calibration
Step 4. Press the
Step 5. Connect the Reflection standard to test port 1 and press the
reflection measurement is performed, and an asterisk appears at the
Step 6. Connect the Reflection standard to test port 2 and press the
reflection measurement is performed, and an asterisk appears at the
and then a check mark [v] also appears on the
Step 7. Press the
Step 8. Connect the Match standard to test port 1 and press the
measurement is performed, and an asterisk appears at the
Step 9. Connect the Match standard to test port 2 and press the
measurement is performed, and an asterisk appears at the
check mark [v] also appears on the
Step 10. Press the
Step 11. Connect the Line 1 standard between test ports 1 and 2 and press the
590
button (1 in Figure A-6).
Thru
button and then a check mark [v] also appears on the
button (3 in Figure A-6).
Reflection
button (4 in Figure A-6).
Match
button (5 in Figure A-6).
Line1
Thru
[ ]Reflection1
[ ]Reflection2
button.
Reflection
[ ]Match1
[ ]Match1
[ ]Match2
[ ]Match2
button.
Match
button (2 in
[ ]Thru 1-2
[ ]Thru
button.
button. The
button.
[ ]Reflection1
button. The
button
[ ]Reflection2
button. The match
button.
button. The reflection
button and then a
button.
[ ]Line1 1-2