Defining standard for each subclass

This section demonstrates the procedure to specify a different OPEN standard for each
frequency band, using subclasses #1 and #2. In this example, the standards created in
"Defining different standard for each frequency band" on page 202 are used.
Procedure
Step 1. Press
.
Step 2. Press
.
Cal Kit
Step 3. Select a calibration kit you want to use. (In this example, use the calibration kit defined in
Step 3 of "Defining different standard for each frequency band" on page 202.)
Step 4. Press
.
Modify Kit
Step 5. Press
Specify CLSs
Step 6. Press
, and select
Sub Class
Step 7. Select the standard type
Step 8. Select #1 (
Open 3G
Step 9. Press
.
Return
Step 10. Then, press
Sub Class
Step 11. Select the standard type
Step 12. Select #2 (
Open 6G
Step 13. Press
.
Return
Chapter 4
Specifying Different Standard for Each Frequency
.
.
Sub Class 1
, press
, make settings for all ports.
Open
Set All
).
, and select
.
Sub Class 2
, press
, make settings for all ports.
Open
Set All
).
Calibration