Measurement Examples
Measuring the Deviation from a Linear Phase
2. Define the Measurement Conditions
The measurement conditions are defined by following the steps described below.
Setup Description
Presetting
Center frequency: 1.09 GHz
Frequency span: 20 MHz
Measurement parameter: S
21
Data format: Expand Phase
Executing the Auto Scale
3. Execute the Calibration
The THRU response calibration is executed.
Setup Description
Executing the THRU response
4. Connect the DUT and Execute the Auto Scale
The DUT is connected again as shown in Figure 15-21 to execute the auto scale.
Setup Description
Executing the auto scale
506
Key Operation
-
OK
-
S21
-
Expand Phase
-
Auto Scale
Key Operation
(A THRU standard is connected instead of a
DUT)
-
-
Calibrate
Response (Thru)
-
Thru
Done
Key Operation
-
Auto Scale
-
Chapter 15