Multi-Trace Configurations - Keysight Technologies FieldFox N9912A104 User Manual

Table of Contents

Advertisement

NA (Network Analyzer) Mode
NA Mode Settings
— Then choose from the following:

Multi-Trace Configurations

In NA Mode you can display multiple traces on the FieldFox screen.
The graphic above shows a x3H configuration. Tr2 is the ACTIVE trace as
indicated by the highlighted Tr2.
Trace Setting Notes
— The Frequency Range, IF BW, Resolution, Average, and Output Power
settings are common for all displayed traces.
— All other trace settings, such as measurement, format, and limit lines, are
applied individually to the ACTIVE trace in the same manner as when a
single trace is present.
5- 4
—Off
(default) No conversion is performed.
—Z Conv (--) or Y Conv (--)
admittance (Y). Then choose from the following for either:
——Auto
The displayed S-parameter is converted to the
appropriate Z or Y parameter: Refl for S11 and S22; Trans for
S21 and S12. When the S-parameter is changed, the
appropriate conversion changes automatically.
——Refl
The displayed S-parameter is converted to Z or Y
reflection, regardless of whether the S-parameter is reflection
(S11 or S22) or transmission (S21 or S12).
——Trans
The displayed S-parameter is converted to Z or Y
transmission, regardless of whether the S-parameter is
reflection (S11 or S22) or transmission (S21 or S12).
1/S
The displayed S-parameter is converted to 1/S-parameter.
Perform conversion for impedance (Z) or