RADIOMETER ABL800 FLEX Operator's Manual page 99

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Calibration setup,
Calibration
This program allows you to schedule automatic calibrations, manual tHb
schedule
calibration and automatic cleaning (the first automatic cleaning also includes
quality control of the Crea A and B electrodes for high creatinine values on the
ABL8x7 FLEX analyzers), and select a calibration to be performed after a sample
measurement.
To edit the schedule for automatic calibrations/cleaning and tHb calibration, do the
following:
Step
Continued
Action
1.
Highlight the desired calibration or cleaning on the screen.
2.
Press the Edit button.
Use the arrow buttons to select start time for calibrations and cleaning
and the intervals between each activity.
3. Installation and setup
24-hour scale
shows the time for
each scheduled
automatic
calibration or
cleaning.
List of activities:
calibrations and
cleaning.
Activity after
each
measurement
Continued on next page
3-33