RADIOMETER ABL800 FLEX Operator's Manual page 122

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

3. Installation and setup
Parameters and input setup,
Parameters
Parameters can be locked/unlocked from the RADIANCE system.
(continued)
To lock or unlock a parameter from the analyzer, do the following:
Step
To edit the parameter setup, do the following.
Step
3-56
Action
1.
Make sure the analyzer is not connected to the RADIANCE system.
Highlight a parameter on the screen, using the scroll facilities.
2.
Press Lock/Unlock. (This button is grayed out if the analyzer is
connected to the RADIANCE system.)
The locked parameter will show red on the parameter bar and will
change the overall analyzer status traffic light on the Analyzer Status
button to red. The parameter value will be absent from the printout.
However, the locked parameter will be calibrated.
3.
To unlock a parameter, highlight it and press Lock/Unlock. The traffic
light on the Analyzer Status button will change from red to a color
corresponding to the analyzer's overall status.
Action
1.
Highlight the desired parameter and press Edit.
Activate (or deactivate) the following check buttons to select (or
2.
deselect) the following functions:
ABL800 FLEX Operator's Manual
Continued
Highlighted
parameter
Continued on next page