11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-12
Continued
No.
Message
210
Calibration
error(s)
present
211
Quality
control
error(s)
present
212
System
message(s)
present
213
Automatic
backup
failed
214
Automatic
backup
succeeded
216
Printer error The printer is out of
217
Replace-
ment:
Interpretation
An error registered on
one or more
parameters during the
last calibration.
One or more errors
registered during last
QC measurment in one
of the installed QC
levels.
One or more systems
errors are present.
An error occurred
during the scheduled
data backup.
The scheduled
automatic backup was
completed successfully.
paper, paper is jammed,
or some other printer
problem has occurred.
The message is used in
the Activity Log to
indicate a performed
replacement.
ABL800 FLEX Operator's Manual
Operator action
Check the Calibration
Status for errors in latest
calibration results for the
given parameter. View
calibration error messages
and take required
corrective action.
Check the Quality Control
Status for errors. View QC
error messages and take
required corrective action.
Check the System
Messages Status for errors.
Take required corrective
action.
- Check Automatic Backup
Setup.
- Check network and
servers used for the
backup.
- Contact your IT engineer.
No action required.
- Check printer paper.
Clear any jam or mount a
roll.
- Power down and restart
the analyzer.
- Contact Radiometer
Service Representative.
No action required.
Continued on next page