ABL800 FLEX Operator's Manual
Replacing inlet gasket unit and inlet probe,
Replacing the
Step
inlet probe
(continued)
Action
Disconnect the tube from the inlet probe and the nipple on the inlet
5.
module.
Connect the new tube to the nipple on the inlet module and to the new
6.
probe.
Then insert the inlet probe with the tube into the groove of the
attachment and seat it properly. Close the clip.
Mount the inlet gasket unit and inlet flaps, then press Restart.
7.
7. Replacements
Continued
7-13