RADIOMETER ABL800 FLEX Operator's Manual page 58

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

2. What Is what
Analyzer status,
Quality Control
Buttons:
status
Run AC
(continued)
ampoule
Result
Reagents status
The Reagents status shows the following:
Buttons:
Replace
Adjust Level
Troubleshoot
Close
NOTICE: The Calibration and Cleaning Met II Solution containers are filled so
that they provide up to 14-days of use in the ABL8x7 FLEX analyzer. This will be
registered on the screen by percentages that are lower than 100 % (i.e. 87%, 75%,
50% for the Calibration Solution 1, Calibration Solution 2 and Cleaning Met II
Solution, respectively). This has been made so that the solution warning function
will not be impaired and you can select the warning values in the usual manner.
2-32
Continued
Press to start a measurement on the highlighted AutoCheck
solution.
For ABL8x7 FLEX analyzers only: When Slot 11 is
highlighted, this button changes to Run High Crea Check to
start a measurement on the Cleaning Met II Solution.
Press to display the last measurement result on the control
solution.
status of the solutions and gas cylinders
When Calibration/Cleaning solutions are expired, the container in question
is marked with a clock above it and the contents are colored orange.
messages referring to solutions and gas cylinders.
Press to enter the screen for replacements of fluids – see
chapter 7: Replacements.
Press to enter the Solution Level Adjustment program to make
adjustments – see chapter 3 in this manual.
Press to view the error interpretation and operator actions –
see the description in chapter 11: Troubleshooting.
Press to exit to the Ready screen.
ABL800 FLEX Operator's Manual