RADIOMETER ABL800 FLEX Operator's Manual page 313

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
628
No
Cleaning
Solution
detected by
Oxi liquid
sensor
during
Refill
630
No Cal 2
Solution
detected by
Oxi liquid
sensor
during
Refill
632
No Cal 1
Solution
detected by
Oxi liquid
sensor
during
Refill
634
No Rinse
Solution
detected by
Oxi liquid
sensor
during
Refill
Gas 1
636
cylinder low
pressure
Gas 2
637
cylinder low
pressure
Interpretation
OXI module was not
filled correctly with
Cleaning Solution.
OXI module was not
filled correctly with Cal
2 Solution.
OXI module was not
filled with Cal 1
Solution.
OXI module was not
filled with Rinse
Solution.
Low pressure level is
detected in Gas 1
cylinder.
Low pressure level is
detected in Gas 2
cylinder.
11. Troubleshooting
Operator action
- Check Cleaning solution
level in the container.
Replace as necessary.
- Perform Fluid transport
troubleshooting procedure.
- Check Cal 2 Solution
level in the container.
Replace as necessary.
- Perform Fluid transport
troubleshooting procedure.
- Check Cal 1 Solution
level in the container.
Replace as necessary.
- Perform Fluid transport
troubleshooting procedure.
- Check Rinse Solution
level in the container.
Replace as necessary.
- Perform Fluid transport
troubleshooting procedure.
Replace the Gas 1 cylinder
within a few days.
Removal condition:
Sufficient pressure in Gas 1
cylinder.
Replace the Gas 2 cylinder
within a few days.
Removal condition:
Sufficient pressure in Gas 2
cylinder.
Continued on next page
11-31