RADIOMETER ABL800 FLEX Operator's Manual page 333

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
805
Adaptive
measuring
mode appli-
ed (decay to
pH/BG/Met/
Oxi)
806
Adaptive
measuring
mode appli-
ed (decay to
all micro
mode)
807
Unautho-
rized
software
808
Calibration
drift not
available
809
Service
Electrode
Updatings
disabled
Interpretation
Due to insufficient or
inhomogeneous
sample, the analyzer
has automatically
chosen an adaptive
sample handling in
order to report as many
of the desired
parameters as possible.
Software option key is
not installed. Analyzer
cannot be used.
A calibration drift
value could not be
calculated, since this is
the first calibration
result after calibration
drift monitoring was
started.
This message may
appear on the first
calibration result after
the analyzer has been
turned on or after the
parameter in question
has been enabled.
A drift value will be
calculated for the next
calibration result.
The message is used in
the Activity Log to
indicate that Service
Electrode Updatings
have been disabled.
11. Troubleshooting
Operator action
- If missing parameters are
needed, repeat the
measurement with
sufficient sample free from
clots and air bubbles.
Contact Radiometer service
representative.
No action required. The
message is informative.
No action required. The
message is for information
only.
Continued on next page
11-51