RADIOMETER ABL800 FLEX Operator's Manual page 321

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
705
AutoCheck
QC
measure-
ment is
repeated
twice
707
Replace-
ment
overdue by
10 %.
Analyzer
locked
708
Corrective
action not
possible due
to empty
AutoCheck
carousel
El/Met
709
module not
expected
OXI module
710
not
expected.
Warning:
711
DysHb high
Interpretation
The scheduled QC
measurement was not
accepted; the
measurement was
repeated twice as
requested in Setup
program: Corrective
Actions.
Replacement is
overdue by 10%
(corrective action
"Lock analyzer" was
selected in Setup
program: Corrective
Actions). When the
analyzer is locked,
scheduled calibrations
are performed, but no
patient samples or QC
measurements are
allowed.
Scheduled AutoCheck
measurement was
requested, but the
carousel was empty.
Software configuration
does not match
analyzer type.
Software configuration
does not match
analyzer type.
Blood sample has a
high level of DysHb.
FCOHb + FMetHb >
5%.
11. Troubleshooting
Operator action
No action required.
- Check Replacement
Status and replace as
required.
- Unlock analyzer in
Miscellaneous Setup
program.
Removal condition: No
replacement pending.
Refill AutoCheck carousel.
Contact Radiometer service
representative.
Removal condition:
Matching software and
hardware configuration.
Contact Radiometer service
representative.
Removal condition:
Matching software and
hardware configuration.
Informative message. No
analyzer error was
detected.
Continued on next page
11-39