RADIOMETER ABL800 FLEX Operator's Manual page 87

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analysis setup,
Setting
Step
reference and
critical limits
(continued)
Reportable
To change or enter a reportable range (must be selected smaller than or equal to the
ranges
measuring range), do the following:
Step
Continued
Action
To remove all current ranges, press the Clear limits button.
3.
4.
To change the value in a range limit, touch and highlight the range.
Then enter the limit, using the screen keyboard and confirm with
Enter.
Repeat for other limits in a similar way.
5.
6.
Press Back when completed.
Action
Scroll to the desired parameter, using the up/down arrows or scroll
1.
bar.
2.
Key in the desired lower limit and confirm with Enter on the keypad.
Key in the upper limit and confirm with Enter on the keypad.
3.
4.
To change the reportable range to the default (primary) setting,
highlight the desired parameter and press Set Default.
3. Installation and setup
Measured
parameters show
their measuring
ranges. Derived
parameters show
"....."
Continued on next page
3-21