RADIOMETER ABL800 FLEX Operator's Manual page 185

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Preparing a control solution,
Before measure-
Step
ment on
QUALICHECK+
control solution
(continued)
Before manual
Do the following:
measurement on
AutoCheck
Step
control solution
Action
3.
2
Place the ampoule in the ampoule opener and break off
the ampoule neck.
1
4.
Place the ampoule fully into the H700 adapter.
Action
1.
Place the ampoule in the H705 Ampoule holder.
2.
Holding the ampoule in its holder between two fingers, shake it
vigorously for at least 15 seconds.
3.
Place the H700 Ampoule adapter over the ampoule and press it down
to open the ampoule.
Continued
5. Quality control
5-5